Calibration and measurement uncertainty in the opposing product report: how does a reviewer check the numbers?

When a defect report says that a shaft is 0.02 mm under the drawing tolerance or a coating 3 microns too thin, whether that proves a defect depends on numbers the report often leaves out: the calibration of the instrument and the uncertainty of the measurement. An FBI Laboratory procedure states that a measurement result is complete only with a quantitative statement of its uncertainty. A reviewer checks whether the opposing expert's numbers meet that test.

Hands hold a micrometer above a small metal part next to a gauge block.

Why does measurement uncertainty matter when goods are said to be out of specification?

Because every measured value is an estimate. If the uncertainty interval around the result reaches across the specification limit, the measurement alone does not show which side of the limit the goods are on.

A measurement gives a value and a range of doubt around it. The United Kingdom Accreditation Service (UKAS) publication M3003, The expression of uncertainty and confidence in measurement, edition 6 of March 2024, describes uncertainty as the doubt attached to a measured value, expressed as limits linked to a level of confidence, and requires the measurand to be defined and a measurement model established before the uncertainty is evaluated.

The FBI Laboratory procedure FTD-124-00, Measurement Uncertainty and Validation, issued for the Firearms/Toolmarks Discipline on February 18, 2022, puts the principle in one line: a measurement result is "only complete when accompanied by a quantitative statement of its uncertainty."

A shaft measured at 0.02 mm under the drawing limit with an expanded uncertainty of 0.03 mm may be inside the tolerance. A coating thickness reading 3 microns below the minimum means little if the gauge was last calibrated two years ago and its calibration uncertainty is not stated. The reviewer asks for the numbers that make the reading meaningful.

How is measurement uncertainty evaluated and stated?

Components are evaluated statistically (Type A) or from other information such as calibration data (Type B), combined into a standard uncertainty and multiplied by a coverage factor, usually k = 2 for about 95% coverage.

UKAS M3003 defines the two kinds of evaluation. Under section 4.1, "If an uncertainty is evaluated by statistical analysis of a series of observations, it is known as a Type A evaluation"; repeatability is the usual Type A component. Under section 5.1, an uncertainty evaluated by non-statistical analysis is a Type B evaluation, and section 5.2 lists among its sources the uncertainties imported from calibration, drift, resolution and environmental conditions.

The components are converted to standard uncertainties and combined with sensitivity coefficients into the combined standard uncertainty (sections 3.31 to 3.37). The expanded uncertainty is the combined standard uncertainty multiplied by a coverage factor (section 3.42). M3003 recommends k = 2 and explains: "This value of k will give a coverage probability of approximately 95 %, assuming a normal distribution." Where that assumption fails, appendices give other coverage factors.

Section 6 sets the form of the report. The result is stated as a coverage interval, y ± U, with a standard statement giving the coverage factor and the coverage probability. Uncertainties are generally reported to two significant figures, the value to the same number of decimal places, and rounding is done only at the end of the calculation.

Laboratories can set a higher confidence level for their own work. The FBI procedure requires reported quantitative results in its discipline to use a confidence level of 99.7% or greater, which corresponds to a coverage factor of 3, and a minimum of ten measurements. A Student-t correction factor can be applied when the Type A component dominates and fewer than 100 measurements were taken. A reviewer therefore reads the coverage factor before comparing two reports: an interval at k = 3 is wider than the same data at k = 2.

What does calibration add, and how do laboratories document it?

Calibration ties the instrument's readings to reference standards and supplies its own uncertainty. Laboratories record the certificate, the interval and the uncertainty it states, and feed that uncertainty into the budget for each measurement.

The FBI procedure requires equipment to be NIST traceable and calibrated at least at the intervals designated by the manufacturer. Calibration uncertainties are taken from the certificates, and the procedure notes that certificates from NIST or from an accredited laboratory traceable to NIST typically report a 95% level (2σ), which must be compensated for, by dividing by 2, in the systematic uncertainty calculation.

The procedure also requires a written uncertainty budget. Examiners list the sources of uncertainty, draw a cause-and-effect diagram whose rules remove duplicate sources, check in a reconciliation step whether existing data already account for a listed source, combine the contributions by root-sum-square, and record the budget on a form that the unit's quality assurance representative keeps.

Repeatability raises a point that even specialists dispute. In a paper presented at the NCSLi 2017 Workshop and Symposium, Craig M. Shakarji and Steven D. Phillips of NIST argued that, for the range of instrument testing and calibration cases they documented, the repeatability of the instrument under test should not be included in the test uncertainty, and wrote that companies "have had to greatly and artificially inflate their accuracy specifications" because of the misunderstanding. A reviewer checks which components the opposing budget contains and whether any was counted twice or left out.

UKAS M3003 warns against padding in the other direction. Inflating inputs or outputs as a safety factor is not consistent with the document, and section 3.48 states that decisions about safety, meaning risk associated with measurement results, can only reliably be made after a realistic uncertainty estimate is obtained.

Have courts refused measurements reported without uncertainty?

Some have, in breath and blood alcohol cases. According to the North Carolina Office of Indigent Defense Services, courts in Washington and Michigan suppressed results reported without an uncertainty or confidence level in 2010 and 2011.

Forensic Resources, a website of the North Carolina Office of Indigent Defense Services, lists four trial court decisions. In People v. Jabrocki (79th District Court, Mason County, Michigan, May 6, 2011) the court held that calculating and reporting an uncertainty budget or error rate is "an essential element of the scientific methodology for analyzing blood alcohol content" by gas chromatography and denied the prosecution's motion to admit the blood test results.

In State v. Fausto (King County District Court, Washington, September 21, 2010) and in City of Kent v. McDaniel (May 4, 2011) the courts ordered breath alcohol measurements suppressed because no measurement of uncertainty was provided. In State v. Weimer (Snohomish County District Court, Washington, March 23, 2010) the court held that allowing the numerical value of blood alcohol tests without stating a confidence level violates Rule of Evidence 403.

The same page states that ISO 17025 is the international standard for testing and calibration laboratories and that its section 5.4.6 requires laboratories to estimate the uncertainty of measurements. The page cites the section by that number without naming the edition.

These are criminal cases decided by trial courts, and they turn on legal limits for alcohol. The reasoning carries to goods disputes where a contract or a standard sets a numerical limit: the closer the result is to the limit, the more the conclusion depends on the uncertainty the report does or does not state.

What does the reviewer ask about the instruments and the measurements?

Which instrument was used, when it was calibrated and by whom, what uncertainty the certificate states, how many readings were taken, under what conditions, and how each source entered the uncertainty budget.

The review lists the following for each measurement on which the opposing conclusion depends, with the page of the report.

The instrument by maker, model and serial number, its resolution and its measuring range compared with the value measured.

The calibration certificate: the laboratory that issued it, the date, the due date, whether calibration was traceable to national standards, and the uncertainty the certificate states.

The number of repeat readings, their spread, and whether outliers were removed and on what rule.

The conditions of measurement: temperature, humidity, fixturing and the condition of the sample, compared with any conditions that the standard or the drawing prescribes.

The uncertainty budget: each component, its type, its distribution, the combination method, the coverage factor and the resulting expanded uncertainty.

Whether the report states the result as a value with its expanded uncertainty and coverage factor, or as a bare number.

How does the reviewer test a pass or fail call close to the limit?

By placing the result with its expanded uncertainty next to the limit. If the interval crosses the limit, the report needs a stated decision rule, such as a guard band, before it can call the goods nonconforming.

Three positions are possible. The whole interval lies outside the limit, and the measurement supports a finding of nonconformity. The whole interval lies inside, and the measurement supports conformity. The interval crosses the limit, and the measurement alone does not decide.

For the third case the report needs a decision rule. UKAS M3003 notes that expanded uncertainty may feed into the calculation of guard bands for a conformity decision and refers to its separate publication LAB 48 on conformity with specifications. A contract, a drawing or a product standard may also state how results near the limit are treated.

The reviewer checks whether the opposing report states any decision rule, whether that rule was agreed or set by the governing document, and whether the expert applied it consistently to all the units tested. A report that calls a part out of tolerance on a reading inside the uncertainty interval, without a stated rule, makes a statement the data do not support.

How do the federal rules apply to measurements without calibration or uncertainty?

Rule 702 asks whether the opinion rests on sufficient facts or data and reliable methods reliably applied. Rule 34 lets a party request the instruments, the calibration records and the raw readings.

Under Rule 702 of the Federal Rules of Evidence, a qualified expert may testify if the proponent shows the court "that it is more likely than not that" the testimony "is based on sufficient facts or data", "is the product of reliable principles and methods", and applies those principles and methods reliably to the facts of the case. A measurement taken with an instrument of unknown calibration, or reported without the uncertainty that decides which side of the limit it falls on, bears on each of those elements.

Rule 34 of the Federal Rules of Civil Procedure lets a party request documents and tangible things in the other party's possession, custody or control. For a measurement dispute the request covers the instrument identification, the calibration certificates, the uncertainty budget, the raw readings and the specimens. The request "must specify a reasonable time, place, and manner for the inspection", and the written response is due within 30 days after service.

The review can be used in two ways: as the basis for questions at the deposition of the opposing expert, or as a rebuttal opinion by a specialist who can explain to the court why a reading inside the uncertainty interval does not show a defect.

What does IAE do when the opposing report's measurements are in doubt?

IAE prepares a review of the opposing product defect report, or an expert report on the client's questions, in up to 5 business days. The format, cost, deadline and specialist are confirmed within one business day.

The review page for product defect reports describes the work: the review reads the other side's defect report against Rule 702 and lists the tests the expert did not run and the conclusions the data do not support. Each point in the review cites the rule or decision behind it and the page of the opposing report.

The client sends the opposing report with its exhibits and describes what to check. For a measurement dispute the useful materials are the drawings and specifications, the raw data and photographs of the set-up, the calibration certificates if they were produced, and the expert's notes. The secretariat confirms receipt the same day, and the client approves the candidate and the CV before work starts. A 48-hour rush is available for cases with tight deadlines.

Every specialist is a member of the public IAE registry and can be verified by Member ID or QR code. The specialist confirms in the document that he or she has no interest in the outcome of the case. A review is evidence that the court weighs together with the other materials, and neither IAE nor the reviewer guarantees the outcome. IAE is not a law firm and does not advise on case strategy.

Sources

Read also

All articles

Expert report review

Ready to discuss your case?

Send the materials: the secretariat will confirm the format, cost, deadline and the specialist within one business day.

Expertise request

Received

Cost and timeline estimate

WorkReview of an opinion
TimelineUp to 5 business days
ExpertIAE registry member